MRI Scanner
Siemens 3T Trio
http://www.ohsu.edu/xd/research/centers-institutes/airc/onprc-3t-instrument.cfm
Website(s)
has URL
has_url
"The Siemens 3T Trio instrument is located on the campus of the Oregon National Primate Research Center (ONPRC; 10 miles west of OHSU's main Marquam Hill campus) in a dedicated 2500 ft2 building that includes animal preparation areas and offices. The radio frequency (RF) transmitter unit contains one 35 kW narrowband (1H) RF transmitter and one 8 kW broadband RF transmitter capable of producing 20‑128 MHz. The instrument is equipped with a SONATA gradient coil set and a CASCADE gradient amplifier (2000V/500A on all three axes). Together they are capable of producing field gradients of ≥40 mT/m, with a slew rate of ≥200 T/(m*s), on all axes.
TECHNICAL INFO:
Field Strength/1H Frequency: 3 Tesla (T)/123 MHz
Clear bore/Working bore: 90 cm/60 cm
Gradient Strength/Slew Rate: ≥40 mT/m/ ≥200 T/(m*s)
Receiver channels: 8
Length/Weight: 2.1m/10 tons"
Resource Description
resource_description
Manufacturer
has manufacturer
has_manufacturer
Siemens AG
3T Trio
Model Number
model number
model_number
Related Technique
Magnetic resonance imaging
related technique
related_technique
magnetic resonance imaging
Location
located_in
Advanced Imaging Research Center
located in
2012-08-15T12:41:11.155-05:00
nvasilevsky (Nicole Vasilevsky)
2013-07-16T15:48:43.061-05:00
workflow state
Published
nvasilevsky (Nicole Vasilevsky)
Instrument
instrument
instrument
processed material
processed material
material_entity
independent_continuant
continuant
entity